Our items are commonly identified and trusted by customers and may fulfill continuously switching economic and social wants of Tablet Hardness Tester Price, Welding Cable Connector, Digital Microhardness Tester, We believe that our warm and professional service will bring you pleasant surprises as well as fortune.
OEM Customized Ultrasonic Thickness Gauge Meter - TM510FNplus Coating thickness gauge – TMTeck Detail:

ery (2)

● 2 Measuring modes: continuous/single
● 2 Shutdown modes: manual/automatic
●Wide measuring range with 5 probes available (next page)
●Direct testing mode and block statistics mode (APPL/BATCH)
●Can connect with printer to out of statistical values
●Dataview to connect with PC with USB 2.0 port
●500 datas can be stored

Main features
Can use various probe (F400, F1, F1/90 °, F10, N1, N400, etc.) measurement;
Three calibration methods: one point calibration, two point calibration, the basic calibration;
Display resolution: 0.1 um (measuring range of less than 100 um)
1μ m (range greater than 100 u m)
Have five statistics, data storage 500
There are two working methods: direct ways and means of group
There are two measurements: continuous measurement and a single measurement
There are two shutdown: manual and automatic shutdown shutdown
Can be set Bound: The gauge of the measured value can be automatic alarm and a number of measurements available on the histogram value analysis;
Deleted features: the gross error and error settings can be deleted;
Printing: Print Measurement measurement, statistics, gauges, histogram
A music tone in the operation carried out at any time tips
A power supply under-voltage direct function
An error function
And printers, computer communications (communications software operating environment for the Window operating system) connectivity.

Technical Specifications:

<td ” width=”284″>Measuring range  <td ” width=”284″>0-1250µ with standard probe F1.N1 (10.000mm    max)<td ” width=”284″>Probes available  <td ” width=”284″>5 probes available for F (ferrous: on steel/iron) and N (non-ferrous metals)<td ” width=”284″>Tolerance  <td ” width=”284″>F1:±(1µ+3%H) N1.±(1.5µ+3% H)H: actual thickness tested<td ” width=”284″>Resolution <td ” width=”284″>Alphanumeric with 4 large digits<td ” width=”284″>Operation language <td ” width=”284″>English<td ” width=”284″>Standards <td ” width=”284″>DIN, ISO, ASTM,BS<td ” width=”284″>Min. measuring area <td ” width=”284″>F1:(standard probe)<td ” width=”284″>Min. curvature  radius <td ” width=”284″>convex:3mm, concave:50mm<td ” width=”284″>Min. substratethickness <td ” width=”284″>type F: 0.5mm, type N: 50mm<td ” width=”284″>Calibration <td ” width=”284″>Zero and foil calibration<td ” width=”284″>Statistics  <td ” width=”284″>Number of measurements, mean, standard deviation,maximum and minimum of 3000readings<td ” width=”284″>Data memory <td ” width=”284″>500  measuring data<td ” width=”284″>Limits <td ” width=”284″>Adjustable with acoustic alarm<td ” width=”284″>Interface <td ” width=”284″>USB 2.0<td ” width=”284″>Operating temperature <td ” width=”284″>0-40℃<td ” width=”284″>power supply <td ” width=”284″>AA size 1.5V<td ” width=”284″>Dimensions <td ” width=”284″>125*67*31mm (main unit)<td ” width=”284″>Weight <td ” width=”284″>345.g

Standard Delivery

·  Main unit 1
·  Probe F1 or N1 1
·  Calibration foils 5
·  Instruction manual 1
·   Cetificate 1
·  Warranty card 1
·  Carrying case 1
Communication cable 1
PC software Dataview 1

Optional accessories
●5 probes for different applications
●Calibration foils in various thickness

ery (1)

Type F probe:

<td ” width=”47″>F5<td ” width=”122″> <td ” width=”47″>1.5<td ” width=”47″>

F7

F7

F40

<td ” width=”47″>

0.5

0.5

2

Probe

F400

F1

F1/90°

F10

Measuring Principle Magnetic method
Measuring range(mm)

0400um

01250 um

010000

Min resolution(mm)

0.1

0.1

10

Tolerance One-point calibration (mm)

±(2%H+0.7)

±(2%H+1)

±(2%H+10)

Two-point calibration (mm) ±(1%H+0.7)

±((1%H+1)

±(1%H+10)

Minimum radius of curvature (convex) (mm)

1

1.5

flat

10

 Minimum measuring area(mm)

F3

Minimum thickness of base material(mm)

0.2

Type N probe

Probe N400

N1

CN02

Measuring Principle

Eddy current method

 Measuring range (mm)

0400

01250

10200

 Min resolution(mm)

0.1

0.1

1

Tolerance  One-point calibration (mm)

±(2%H+0.7)

±(2%H+1.5)

±(2%H+1)

 Two-point calibration (mm) ±(1%H+0.7) ±1%H+1.5

——–

Minimum radius of curvature (convex) (mm)

1.5

3

flat

Minimum measuring area(mm)

F4

F5

F7

Minimum thickness of base material(mm)

0.3

0.3

0


Product detail pictures:

OEM Customized Ultrasonic Thickness Gauge Meter - TM510FNplus Coating thickness gauge – TMTeck detail pictures


Related Product Guide:

Our eternal pursuits are the attitude of "regard the market, regard the custom, regard the science" as well as the theory of "quality the basic, believe in the 1st and management the advanced" for OEM Customized Ultrasonic Thickness Gauge Meter - TM510FNplus Coating thickness gauge – TMTeck , The product will supply to all over the world, such as: Colombia, Brunei, Karachi, All styles appear on our website are for customizing. We meet up to personal requirements with all products of your very own styles. Our concept is to help presenting the confidence of each buyers with the offering of our most sincere service, and the right product.
  • In general, we are satisfied with all aspects, cheap, high-quality, fast delivery and good procuct style, we will have follow-up cooperation!
    5 Stars By Janet from UK - 2017.08.28 16:02
    The after-sale warranty service is timely and thoughtful, encounter problems can be resolved very quickly, we feel reliable and secure.
    5 Stars By Cindy from Belgium - 2018.09.23 17:37
    Write your message here and send it to us