INSTRUCTION FOR USE
1. SCOPE
| TYPE | MEASUREMENT RANGE ( GAUSS) | ACCURACY | SIZE (mm) |
| MG-2 | 20-0-20 | +/-10% | 60mm DIA.X23mm |
2. STORAGE
THIS METER CAN BE USED IN THE AMBINET OF -10 ‘C TO +50 ‘C AND RELATIVE HUMIDITY 80%
3. USING METHOD
DURING MEASUREMENT, IT SHOULD BE PLACE ON THE SURFACE OF PART BEING TESTING TIGHTLY WITH INDICATED ARROW, AND MOVED WITH TWO SIDE DIRECTION OF THE INSTRUMENT. AT THE MOVING, THE MAXIMUM VALUE TO BE ACQUIRED IS THE RESIDUAL FIELD STRENGTH
4. CAUTION
BECAUSE THIS IS VERY SENSITIVE MAGNETIZING FIELD STRENGH INDICATOR, IT WILL BE INDUCED BY THE EARTH FIELD. THEREFOR, AT THE MEASUREMENT, THE INSTRUMENT MOVEMENT MUST BE TAKEN THE ORIENTATION FROM EAST TO WEST OR VICE VERSA. IF IT CAN NOT DO THIS WAY, AND THE MOVE IS NOT THE ORIENTATION OF EAST TO WEST, THE MEASUREMENT VALUE SHOULD BE CONSIDERED TO REDUCE THE STRENGTH FROM EARTH FIELD.
5. UNIT CONVERSION
1 GAUSS = 1 X 10 MILLI-TESLA
Product detail pictures:
Related Product Guide:
Our well-equipped facilities and exceptional good quality manage throughout all stages of production enables us to guarantee total shopper satisfaction for 2021 Good Quality Phased Array Flaw Detector - MG-2 MAGNETIZING FIELD STRENGTH INDICATOR – TMTeck , The product will supply to all over the world, such as: Cyprus, Cancun, London, Our domestic website's generated over 50, 000 purchasing orders every year and quite successful for internet shopping in Japan. We would be happy to have an opportunity to do business with your company. Looking forward to receiving your message !